Problem:High background
Possible causes 1:Reagents are not at room temperature
Solutions 1:Allow all reagents to reach room temperature (18-25℃) before use.
Possible causes 2:Incomplete washing
Solutions 2:Ensure that the washing apparatus is working correctly. Ensure adequate time of soaking.
Possible causes 3:Inadequate aspiration of wells
Solutions 3:Wells should be fully dry after aspiration.